Combined analysis:

Analysis ID AN002713
Analysis type MS
Chromatography type Reversed phase
Chromatography system Agilent 6550
Column Waters Acquity BEH C18 (100 x 2mm,1.7um)
MS Type EI
MS instrument type QTOF
MS instrument name Agilent 6550 QTOF
Ion Mode POSITIVE
Units peak area

MS:

MS ID:MS002510
Analysis ID:AN002713
Instrument Name:Agilent 6550 QTOF
Instrument Type:QTOF
MS Type:EI
MS Comments:Full scan MS data from 50 to 1700 m/z with a scan frequency of 6 Hz was collected both in positive (ESI+) and negative (ESI-) electrospray ionization modes. The following electrospray ionization parameters were used: gas temperature, 200 °C; drying gas, 14 l min-1; nebulizer, 60 psi; sheath gas temperature, 350 °C; sheath gas flow, 11 l min-1. Urine samples were randomly injected in the chromatographic system in order to avoid intra-batch variability, as well as to enhance quality and reproducibility. QC sample was analyzed every 7 injections to monitor and correct changes in the instrument response. QC sample was also repeatedly analysed under auto MS/MS and All-ion (MSE) fragmentation modes to provide useful information of fragment ions for identification purposes. Sample stability and analytical drift were investigated through IS intensities. Blank analysis was performed at the end of the sequence and used to identify artefacts from sampling, preparation of samples and analysis.
Ion Mode:POSITIVE