Return to study ST001733 main page

MB Sample ID: SA162649

Local Sample ID:POL_19
Subject ID:SU001810
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002821 AN002822
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Agilent 1200 Agilent 1200
Column Discovery HS C18 (150 x 2.1mm,3.0um) Discovery HS C18 (150 x 2.1mm,3.0um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Agilent 6520 QTOF Agilent 6520 QTOF
Ion Mode NEGATIVE POSITIVE
Units peak area peak area

MS:

MS ID:MS002615
Analysis ID:AN002821
Instrument Name:Agilent 6520 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:The capillary voltage was set at 4,000 V. The drying gas flow rate was 10.5 L/min at 330 ℃ and gas nebulizer at 52 psi; fragmentor voltage was 175 V; skimmer and octopole radio frequency voltage (OCT RF Vpp) were set to 65 and 750 V. Data were collected in the centroid mode at a scan rate of 1.2 spectra per second. Mass spectrometry detection was performed in full scan from 100 to 1200 m/z. The reference m/z ions were TFA NH4 (119.0363) and HP-0921 (966.0007). These masses were continuously infused into the system to allow constant mass correction. Samples were analysed in separate runs. Acquired data were cleaned of background noises and unrelated ions using MassHunter Profinder (B.06.00, Agilent Technologies) software. “Molecular feature extraction” and “Find by ion” algorithms were applied to reduce the size and complexity of data, and to improve the reliability in finding the features. 698 chemical signals were obtained. Then, data was filtered, and only those features detected in >50% in QCs and with a Relative Standard Deviation (RSD) <30% in Qcs were kept, resulting in 429 signals.
Ion Mode:NEGATIVE
  
MS ID:MS002616
Analysis ID:AN002822
Instrument Name:Agilent 6520 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:The capillary voltage was set at 3,500 V. The drying gas flow rate was 10.5 L/min at 330 ℃ and gas nebulizer at 52 psi; fragmentor voltage was 175 V; skimmer and octopole radio frequency voltage (OCT RF Vpp) were set to 65 and 750 V. Data were collected in the centroid mode at a scan rate of 1.2 spectra per second. Mass spectrometry detection was performed in full scan from 100 to 1200 m/z. The reference m/z ions were purine (121.0508) and HP-0921 (922.0097). These masses were continuously infused into the system to allow constant mass correction. Samples were analysed in separate runs. Acquired data were cleaned of background noises and unrelated ions using MassHunter Profinder (B.06.00, Agilent Technologies) software. “Molecular feature extraction” and “Find by ion” algorithms were applied to reduce the size and complexity of data, and to improve the reliability in finding the features. 1654 chemical signals were obtained. Then, data was filtered, and only those features detected in >50% in QCs and with a Relative Standard Deviation (RSD) <30% in Qcs were kept, resulting in 535 signals.
Ion Mode:POSITIVE
  logo