MB Sample ID: SA162664

Local Sample ID:POL-36
Subject ID:SU001811
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002823
Analysis type MS
Chromatography type HILIC
Chromatography system Agilent 1260
Column Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um)
MS Type ESI
MS instrument type Triple quadrupole
MS instrument name Agilent 6470 QQQ
Ion Mode NEGATIVE
Units ug/mg

MS:

MS ID:MS002617
Analysis ID:AN002823
Instrument Name:Agilent 6470 QQQ
Instrument Type:Triple quadrupole
MS Type:ESI
MS Comments:The MS conditions were: 5500 V of capillary voltage in positive ESI mode, a nebulizer gas flow rate of 11.0 L/min; a source temperature of 250 °C; and a source pressure of 60 psi. The sample tray temperature was maintained at 4 °C. Each transition was optimized adjusting the fragmentor and collision energy voltages. Data were acquired in MassHunter Workstation B.05.00 (Agilent Technologies), and re-processed using MassHunter QQQ Quantitative Analysis B.08.00 (Agilent) where peak areas were integrated. Concentration of metabolites were calculated using calibration curves with the standard addition method.
Ion Mode:NEGATIVE