MB Sample ID: SA164970

Local Sample ID:S7R3
Subject ID:SU001853
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002883
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Dionex Ultimate 3000 RS
Column Thermo Accucore HILIC (100 x 2.1mm,2.6um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode POSITIVE
Units arbitray unit

MS:

MS ID:MS002676
Analysis ID:AN002883
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The MS acquisition parameters were as follows: spray voltage, 3.5 kV; capillary temperature, 320 °C; sheath gas flow rate, 45; auxiliary gas flow rate, 25; heater temperature 350 °C; AGC, 3×106; maximum injection time, 200 ms; mass scan range, 50–750; full MS resolution, 70,000 FWHM at m/z 200; spectrum data type, profile. The raw LC-HRMS files for the same study were processed in one batch using the label-free differential analysis software (SIEVE 2.2, Thermo Scientific), in which the ChromAlign algorithm was used. The key parameters for the feature extraction were as follows: signal to background noise,>3; Mzstep accuracy, 10 ppm, minimum peak intensity 200,000; minimum peak scan points, 5; minimum isotopes,1.
Ion Mode:POSITIVE