MB Sample ID: SA205251

Local Sample ID:sample1-2
Subject ID:SU002221
Subject Type:Synthetic sample

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003497
Analysis type MS
Chromatography type CE
Chromatography system ZipChip (908 Devices)
Column HS chip
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Exactive Plus Orbitrap
Ion Mode POSITIVE
Units Micromolars

MS:

MS ID:MS003257
Analysis ID:AN003497
Instrument Name:Thermo Exactive Plus Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:All experiments were performed in positive ionization mode in the 50-500 m/z range at a mass resolution setting of 17,500. The capillary temperature was set to 200 °C and the sheath gas flow rate was 2 psi. The automatic gain control (AGC) target value was set to 3E6 and the maximum injection time was 20 ms. Data were acquired using Xcalibur 3.0 (Thermo Scientific) and were imported to Skyline software14 for peak picking and integration. The peak picking procedure used the analyte accurate m/z and migration time. Peak areas obtained from Skyline were exported as spreadsheets for further analysis. Quantitation was performed with the analyte peak areas relative to the peak area of one of the three isotopically labeled internal standards (13C6 arginine, 13C methionine D3 and 13C phenylalanine) chosen based on migration time similarities.
Ion Mode:POSITIVE