Return to study ST002911 main page
MB Sample ID: SA316308
Local Sample ID: | 1_UT6 |
Subject ID: | SU003024 |
Subject Type: | Mammal |
Subject Species: | Mus musculus |
Taxonomy ID: | 10090 |
Age Or Age Range: | Seven- to eight-week-old female BALB/c mice |
Gender: | Female |
Animal Animal Supplier: | Jackson Laboratories |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004780 | AN004781 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | Reversed phase | Reversed phase |
Chromatography system | Agilent 1290 Infinity II | Agilent 1290 Infinity II |
Column | Agilent InfinityLab Poroshell 120 EC-C18 (100 x 3mm,2.7um) | Agilent InfinityLab Poroshell 120 EC-C18 (100 x 3mm,2.7um) |
MS Type | ESI | ESI |
MS instrument type | QTOF | QTOF |
MS instrument name | Agilent 6545 QTOF | Agilent 6545 QTOF |
Ion Mode | POSITIVE | NEGATIVE |
Units | AREA | AREA |
MS:
MS ID: | MS004526 |
Analysis ID: | AN004780 |
Instrument Name: | Agilent 6545 QTOF |
Instrument Type: | QTOF |
MS Type: | ESI |
MS Comments: | 150 V fragmentor, 65 V skimmer, 3500 V capillary voltage, 750 V octopole radio frequency voltage, 10 L/min nebulizer gas flow, 200 °C gas temperature, 50 psi nebulizer gas pressure, 12 L/min sheath gas flow, and 300 °C sheath gas temperature. Data were collected in positive and negative ESI modes in separate runs, operated in full scan mode from 40 to 1700 m/z with a scan rate of 3 spectra/s. A solution consisting of two reference mass compounds was infused throughout the whole analysis: purine (C5H4N4) at m/z 121.0509 and HP-0921 (C18H18O6N3P3F24) at m/z 922.0098. These masses were continuously infused into the system through an Agilent 1260 Iso Pump at a 1 mL/min (split ratio 1:100) to provide a constant mass correction. |
Ion Mode: | POSITIVE |
MS ID: | MS004527 |
Analysis ID: | AN004781 |
Instrument Name: | Agilent 6545 QTOF |
Instrument Type: | QTOF |
MS Type: | ESI |
MS Comments: | 150 V fragmentor, 65 V skimmer, 3500 V capillary voltage, 750 V octopole radio frequency voltage, 10 L/min nebulizer gas flow, 200 °C gas temperature, 50 psi nebulizer gas pressure, 12 L/min sheath gas flow, and 300 °C sheath gas temperature. Data were collected in positive and negative ESI modes in separate runs, operated in full scan mode from 40 to 1700 m/z with a scan rate of 3 spectra/s. A solution consisting of two reference mass compounds was infused throughout the whole analysis: purine (C5H4N4) at m/z 119.0363 and HP-0921 (C18H18O6N3P3F24) at m/z 980.0163 (HP-0921 + acetate). These masses were continuously infused into the system through an Agilent 1260 Iso Pump at a 1 mL/min (split ratio 1:100) to provide a constant mass correction. |
Ion Mode: | NEGATIVE |