Analysis ID AN005001
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Vanquish
Column Phenomenex Kinetex XB-C18 (150 x 2.1mm, 2.6um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode NEGATIVE
Units normalized AUC

MS:

MS ID:MS004741
Analysis ID:AN005001
Instrument Name:Thermo Q Exactive HF hybrid Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The Vanquish UHPLC system was coupled to a Q ExactiveTM HF Quadrupole-OrbitrapTM high-resolution mass spectrometer (Thermo Fisher Scientific, Austria), equipped with a HESI source for negative ionization to perform the mass spectrometric analysis. The MS scan range was 250-700 m/z with a resolution of 60,000 (at m/z 200) on the MS1 level. A Top 2 method was applied for fragmentation (HCD 24 normalized collision energy), preferable 33 m/z values specific for well-known eicosanoids and precursor molecules from an inclusion list. The resulting fragments were analysed on the MS2 level at a resolution of 15,000 (at m/z 200). Operating in negative ionization mode, a spray voltage of 3.5 kV and a capillary temperature of 253°C were applied. Sheath gas was set to 46 and the auxiliary gas to 10 (arbitrary units).
Ion Mode:NEGATIVE