Return to study ST003172 main page

MB Sample ID: SA343135

Local Sample ID:20230216-M-GL-LJ-NEG7-1
Subject ID:SU003291
Subject Type:Plant
Subject Species:Chili Pepper (Capsicum Chinense)

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN005206
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters Acquity H-Class
Column Waters ACQUITY UPLC HSS T3 (150 x 2.1mm, 1.8um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Orbitrap Exploris™ 240 Mass Spectrometer
Ion Mode UNSPECIFIED
Units peak area

MS:

MS ID:MS004939
Analysis ID:AN005206
Instrument Name:Orbitrap Exploris™ 240 Mass Spectrometer
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The metabolome profiling was carried out using untargeted metabolomics based on liquid chromatography coupled with mass spectrometry (LC-MS), with data collected in positive and negative modes respectively to improve coverage. The samples were filtered through a 0.22 µm membrane and transferred into the lining tube of a sampling vial. Subsequent centrifugation was carried out at 12000 rcf and 4°C for 10 minutes. The processed samples were then analyzed using Thermo Scientific Orbitrap Exploris™ 240 (Thermo Fisher Scientific, USA). Chromatographic separation was achieved on a T3 C18 (1.7 µm, 2.1 mm × 150 mm column, USA) maintained at 40°C. The mobile phase consisted of A: 1% formic acid in water and B: 1% formic acid in acetonitrile, with a flow rate of 300 µL/min. A 3 µL sample was injected at an autosampler temperature of 10°C. The elution gradient was set as follows: 0-2.5 min, 3-10% B; 2.5-6 min, 10-44% B; 6-14 min, 44-80% B; 14-20 min, 80-95% B; 20-23 min, 95% B; 23-23.1 min, 95-3% B; 23.1-28 min, 3% B. MS was performed using both positive and negative ion scans, with a precursor ion scan mode. The auxiliary gas heater temperature was set at 350°C, and the ion transfer tube temperature was also maintained at 350°C. The sheath gas flow rate and auxiliary gas flow rate were set to 35 arb and 15 arb, respectively. The voltages were set to 3.5 KV for the positive spectrum and 3.2 KV for the negative spectrum. For MS1, the scan resolution was 60000, with a scan range of 80-1200. For MS2, the scan resolution was 15000, with a stepped collision energy of 20, 40, and 60 eV. Metabolite identification and quantification were performed using the Compound Discoverer software 3.3 (Thermo Fisher Scientific, USA).
Ion Mode:UNSPECIFIED
  logo