Summary of Study ST001873
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001182. The data can be accessed directly via it's Project DOI: 10.21228/M8XD6P This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST001873 |
Study Title | Metabolomics analysis of multiple samples on AB 5600-Part 1 |
Study Type | Metabolomics |
Study Summary | Metabolomics analysis of multiple samples from human, trying to annotate the metabolites in them. AB SCIEX 5600+ was used for metabolomics detection. |
Institute | Dalian Institute Of Chemical Physics |
Laboratory | Laboratory of High Resolution Separation/Analysis and Metabonomics |
Last Name | Zheng |
First Name | Fujian |
Address | No. 457 Zhongshan Road, Shahekou District, Dalian, Liaoning Province, China |
zhengfj@dicp.ac.cn | |
Phone | 18698730176 |
Submit Date | 2021-06-18 |
Raw Data Available | Yes |
Raw Data File Type(s) | wiff |
Analysis Type Detail | LC-MS |
Release Date | 2021-07-24 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003035 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Waters Acquity |
Column | Waters Acquity BEH C8 (100 x 2.1mm,1.7um) |
MS Type | ESI |
MS instrument type | QTOF |
MS instrument name | ABI Sciex 5600+ TripleTOF |
Ion Mode | POSITIVE |
Units | peak area |
MS:
MS ID: | MS002822 |
Analysis ID: | AN003035 |
Instrument Name: | ABI Sciex 5600+ TripleTOF |
Instrument Type: | QTOF |
MS Type: | ESI |
MS Comments: | The spray voltage was 5.5 kV and the source temperature was 550 °C, Curtain gas, gas 1 and gas 2 were set at 35, 55, and 55 psi, respectively. The data was processed by XCMS, MS-DIAL, MZmine 2 and MetEx. |
Ion Mode: | POSITIVE |