Summary of Study ST003310

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR002057. The data can be accessed directly via it's Project DOI: 10.21228/M8QN7M This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST003310
Study TitleDetermination the concentration of endo-borneol from needle leachates of Pinus koraiensis flushed by precipitation via indoor simulated leaching)
Study SummaryThe precipitation from April to October (2016-2021) was provided by local weather burea in Benxi. The monthly precipitation and duration in Benxi is classified into high, medium, and low level. Under each precipitation, the concentration of endo-borneol is quantified by indoor simulated leaching. The simulated leachates were extracted by ethyl acetate and then analyzed by GC/MSD. As a result, the peak area of endo-borneol ranges from 2006.257 to 27106.75 under tested precipitation.
Institute
Yunnan University
DepartmentSchool of Agriculture
LaboratoryKey Laboratory of Agro-Biodiversity and Pest Management of Education Ministry of China
Last NameXing-Yu
First NameJi
AddressRoad Fengyuan, District Panlong, Kunming, Yunnan, China
Emailxingyu.ji@hotmail.com
Phone+86-13354989110
Submit Date2024-06-05
Raw Data AvailableYes
Raw Data File Type(s)d
Analysis Type DetailGC-MS
Release Date2024-07-15
Release Version1
Ji Xing-Yu Ji Xing-Yu
https://dx.doi.org/10.21228/M8QN7M
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN005422
Analysis type MS
Chromatography type GC
Chromatography system Agilent 8860 GC
Column Agilent HP5-MS (30m x 0.25mm, 0.25 um)
MS Type EI
MS instrument type GC/MSD
MS instrument name Agilent 5977B
Ion Mode POSITIVE
Units Peak area

MS:

MS ID:MS005148
Analysis ID:AN005422
Instrument Name:Agilent 5977B
Instrument Type:GC/MSD
MS Type:EI
MS Comments:An electron ionization system with ionization energy of 7 eV was used. Helium gas (99.999%) was used as the carrier gas at constant flow rate 1 ml/min and an injection volume of 1 ul was employed. The ion-source temperature was 230 °C. The scan interval of mass spectra was 3.1 units/s and fragments from 35 to 270 da. Afterwards, Agilent MassHunter Workstation Quantitative Analysis B.07.01 were used as analytical software. The mass spectrum of the individual peak was inspected manually to confirm the peak had the correct m/z diagnostic fragments, matched with the search results from NIST MS Search 2.2 library. For endo-borneol, the target ion is m/z 95, and qualifier ions are m/z 110 and m/z 67.
Ion Mode:POSITIVE
  logo