Summary of Study ST000324

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000261. The data can be accessed directly via it's Project DOI: 10.21228/M8230J This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST000324
Study TitleProgesterone level effects on primary metabolites in uterus, blood, and ovaries (Part 3:Ovaries)
Study SummaryIn this experiment a hormonal protocol was applied to control follicle growth to yield larger or smaller preovulatory follicle and CLs and consequently different circulating Progesterone (P4) concentrations during early diestrus. The two different animal's group are: high or low progesterone levels. The effects of these progesterone levels was tested in the ovaries (follicle fluid) of the cow.
Institute
University of California, Davis
DepartmentGenome and Biomedical Sciences Facility
LaboratoryWCMC Metabolomics Core
Last NameFiehn
First NameOliver
Address1315 Genome and Biomedical Sciences Facility, 451 Health Sciences Drive, Davis, CA 95616
Emailofiehn@ucdavis.edu
Phone(530) 754-8258
Submit Date2015-03-16
Num Groups1
Total Subjects20
Raw Data AvailableYes
Raw Data File Type(s)d
Analysis Type DetailLC-MS
Release Date2016-01-27
Release Version1
Oliver Fiehn Oliver Fiehn
https://dx.doi.org/10.21228/M8230J
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Project:

Project ID:PR000261
Project DOI:doi: 10.21228/M8230J
Project Title:Progesterone level effects on primary metabolites in uterus, blood, and ovaries
Project Summary:In this experiment a hormonal protocol was applied to control follicle growth to yield larger or smaller preovulatory follicle and CLs and consequently different circulating Progesterone (P4) concentrations during early diestrus. The two different animal's group are: high or low progesterone levels. The effects of these progesterone levels was tested in the uterus, blood, and ovaries of the cow.
Institute:University of California, Davis
Department:Genome and Biomedical Sciences Facility
Laboratory:WCMC Metabolomics Core
Last Name:Fiehn
First Name:Oliver
Address:1315 Genome and Biomedical Sciences Facility, 451 Health Sciences Drive, Davis, CA 95616
Email:ofiehn@ucdavis.edu
Phone:(530) 754-8258
Funding Source:U24DK097154

Subject:

Subject ID:SU000344
Subject Type:Animal
Subject Species:Bos taurus
Taxonomy ID:9913
Species Group:Mammal

Factors:

Subject type: Animal; Subject species: Bos taurus (Factor headings shown in green)

mb_sample_id local_sample_id Organ
SA014652198Follicle Fluid
SA014653199Follicle Fluid
SA014654200Follicle Fluid
SA014655197Follicle Fluid
SA014656196Follicle Fluid
SA014657192Follicle Fluid
SA014658193Follicle Fluid
SA014659194Follicle Fluid
SA014660195Follicle Fluid
SA014661191Follicle Fluid
Showing results 1 to 10 of 10

Collection:

Collection ID:CO000338
Collection Summary:Follicular Fluid: Follicular Fluid were obtained by postmortem aspiration of the ovarian dominant follicle and kept at 80ÂșC. There was no inactivation or attenuation because samples are not biohazard.
Collection Protocol Filename:StudyDesign-MarioBinelli-101414.pdf
Sample Type:Follicular fluid

Treatment:

Treatment ID:TR000358
Treatment Summary:There was no specific treatment. A hormonal protocol was applied to control follicle growth to yield larger or smaller preovulatory follicle and CLs and consequently different circulating Progesterone (P4) concentrations during early diestrus. The two different animal's group are: high or low progesterone levels.
Treatment Protocol Filename:StudyDesign-MarioBinelli-101414.pdf

Sample Preparation:

Sampleprep ID:SP000352
Sampleprep Summary:-
Sampleprep Protocol Filename:SOP_Extraction_of_Mammalian_Tissue_Samples.pdf

Combined analysis:

Analysis ID AN000516 AN000517
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Unspecified Unspecified
Column Waters Acquity CSH C18 (100 x 2.1mm,1.7um) Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Agilent 6530 QTOF Agilent 6550 QTOF
Ion Mode POSITIVE NEGATIVE
Units Peak height Peak height

Chromatography:

Chromatography ID:CH000362
Chromatography Summary:UPLC
Methods Filename:Data_Dictionary_Fiehn_laboratory_CSH_QTOF_lipidomics_05-29-2014.pdf
Instrument Name:Unspecified
Column Name:Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
Column Pressure:450-850 bar
Column Temperature:65 C
Flow Gradient:15% B to 99% B
Flow Rate:0.6 mL/min
Internal Standard:See data dictionary
Retention Time:See data dictionary
Sample Injection:1.67 uL
Solvent A:60% acetonitrile/40% water; 10mM formic acid; 10mM ammonium formate
Solvent B:90% isopropanol/10% acetonitrile; 10mM formic acid; 10mM ammonium formate
Analytical Time:13 min
Capillary Voltage:3500
Time Program:15 min
Weak Wash Solvent Name:Isopropanol
Weak Wash Volume:5 seconds
Strong Wash Solvent Name:Isopropanol
Target Sample Temperature:Autosampler temp 4 C
Randomization Order:Excel
Chromatography Type:Reversed phase

MS:

MS ID:MS000452
Analysis ID:AN000516
Instrument Name:Agilent 6530 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:Lipidomics runs are MS1 data; Identifications made from pooled MS/MS runs
Ion Mode:POSITIVE
Capillary Voltage:3500
Collision Gas:Nitrogen
Dry Gas Flow:8 l/min
Dry Gas Temp:325
Fragment Voltage:120
Fragmentation Method:Auto MS/MS
Ion Source Temperature:325
Ion Spray Voltage:1000
Ionization:Pos
Precursor Type:Intact Molecule
Reagent Gas:Nitrogen
Source Temperature:325 C
Desolvation Gas Flow:11 l/min
Desolvation Temperature:350 C
Nebulizer:35 psig
Octpole Voltage:750
Resolution Setting:Extended Dynamic Range
Scan Range Moverz:60-1700 Da
Scanning Cycle:2 Hz
Scanning Range:60-1700 Da
Skimmer Voltage:65
  
MS ID:MS000453
Analysis ID:AN000517
Instrument Name:Agilent 6550 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:Lipidomics runs are MS1 data; Identifications made from pooled MS/MS runs
Ion Mode:NEGATIVE
Capillary Voltage:3500
Collision Gas:Nitrogen
Dry Gas Flow:8 l/min
Dry Gas Temp:325
Fragment Voltage:120
Fragmentation Method:Auto MS/MS
Ion Source Temperature:325
Ion Spray Voltage:1000
Ionization:Pos
Precursor Type:Intact Molecule
Reagent Gas:Nitrogen
Source Temperature:325 C
Desolvation Gas Flow:11 l/min
Desolvation Temperature:350 C
Nebulizer:35 psig
Octpole Voltage:750
Resolution Setting:Extended Dynamic Range
Scan Range Moverz:60-1700 Da
Scanning Cycle:2 Hz
Scanning Range:60-1700 Da
Skimmer Voltage:65