Summary of Study ST000617
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000451. The data can be accessed directly via it's Project DOI: 10.21228/M8RP6W This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
Study ID | ST000617 |
Study Title | Validation of the application of targeted metabolomic appraoch in the diagnosis of CFS |
Study Type | Plasma metabolomic profiling |
Study Summary | This study was to validate the utility of the developed targeted metabolomic method in the diagnosis of chronic fatigue syndrome (CFS). Clinical validation consisted of a cohort of 20 male CFS (53 ± 2.8 years old, mean ± SEM, range 21-67 y) and 18 male controls (53 ± 3.5 years old, mean ± SEM, range 23-69 y), who were enrolled in a previous study (Naviaux et al. 2016). These plasma samples were stored in -80 ºC for about 1.5 years and reanalyzed on a different LC-MS/MS system by a different investigator. |
Institute | University of California, San Diego |
Department | Department of Medicine |
Laboratory | The Mitochondrial and Metabolic Disease Center |
Last Name | Naviaux |
First Name | Robert |
Address | 214 Dikinson Street, CTF-C102, San Diego, CA, 92103 |
rnaviaux@ucsd.edu | |
Phone | 619-993-2904 |
Submit Date | 2017-05-17 |
Num Groups | 2 |
Total Subjects | 38 |
Num Males | 38 |
Study Comments | These plasma samples were stored in -80 ºC for about 1.5 years and reanalyzed on a different LC-MS/MS system by a different investigator. |
Raw Data File Type(s) | wiff |
Analysis Type Detail | LC-MS |
Release Date | 2017-06-17 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Project:
Project ID: | PR000451 |
Project DOI: | doi: 10.21228/M8RP6W |
Project Title: | Clinical validation for the utility of targeted metabolomic method in chronic fatigue syndrome |
Project Type: | Plasma samples |
Project Summary: | Chronic fatigue syndrome (CFS) is a complex, multisystem disease that has affected more than 2 million people in the United States. No biochemical diagnostic lab test is currently available.As a practical test of the reproducibility of our new metabolomic method, we reanalyzed 40 samples from an earlier CFS study |
Institute: | University of California, San Diego |
Department: | Department of Medicine |
Last Name: | Naviaux |
First Name: | Robert |
Address: | 214 Dickinson St., Bldg CTF, Rm C102, San Diego, CA, 92103, USA |
Email: | naviaux@ucsd.edu |
Phone: | 619-543-2904 |
Subject:
Subject ID: | SU000640 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Age Or Age Range: | 53 ± 3.5 (23-69 yr) |
Gender: | Male |
Human Ethnicity: | White/non-Hispanic |
Species Group: | Human |
Factors:
Subject type: Human; Subject species: Homo sapiens (Factor headings shown in green)
mb_sample_id | local_sample_id | Treatment |
---|---|---|
SA034196 | 211_CFS | CFS |
SA034197 | 213_CFS | CFS |
SA034198 | 214_CFS | CFS |
SA034199 | 210_CFS | CFS |
SA034200 | 202_CFS | CFS |
SA034201 | 206_CFS | CFS |
SA034202 | 201_CFS | CFS |
SA034203 | 215_CFS | CFS |
SA034204 | 207_CFS | CFS |
SA034205 | 218_CFS | CFS |
SA034206 | 231_CFS | CFS |
SA034207 | 228_CFS | CFS |
SA034208 | 208_CFS | CFS |
SA034209 | 224_CFS | CFS |
SA034210 | 225_CFS | CFS |
SA034211 | 221_CFS | CFS |
SA034212 | 223_CFS | CFS |
SA034213 | 205_CFS | CFS |
SA034214 | 219_CFS | CFS |
SA034215 | 209_CFS | CFS |
SA034216 | M46MP_010 | Control |
SA034217 | M23CV_036 | Control |
SA034218 | M26JC_040 | Control |
SA034219 | M37AF_034 | Control |
SA034220 | M54MD_004 | Control |
SA034221 | M66CC_039 | Control |
SA034222 | M65WA_030 | Control |
SA034223 | M47WA_009 | Control |
SA034224 | M27RW_001 | Control |
SA034225 | M58JG_031 | Control |
SA034226 | M64EG_032 | Control |
SA034227 | M67NN_033 | Control |
SA034228 | M69JB_037 | Control |
SA034229 | M63RD_029 | Control |
SA034230 | M62JR_041 | Control |
SA034231 | M58RN_007 | Control |
SA034232 | M59JM_035 | Control |
SA034233 | M56MF_003 | Control |
Showing results 1 to 38 of 38 |
Collection:
Collection ID: | CO000634 |
Collection Summary: | Blood samples from 18 male subjects with chronic fatigue syndrome and 20 age and sex matched controls were collected. Plasma was prepared according to the standard protocol. The metabolites were extracted from |
Sample Type: | Blood |
Collection Method: | Lithium-heparin vacutainer tubes |
Storage Conditions: | -80 C |
Collection Vials: | Lithium-heparin vacutainer tubes |
Collection Tube Temp: | -80 C |
Blood Serum Or Plasma: | Plasma |
Treatment:
Treatment ID: | TR000654 |
Treatment Summary: | No treatment. There are only two groups: control and disease group (Chronic fatigue syndrome) |
Sample Preparation:
Sampleprep ID: | SP000647 |
Sampleprep Summary: | Metabolites from plasma were extracted using pre-chilled methanol/acetonitrile (50:50) |
Extraction Method: | Typically, 90 μl of plasma was thawed on ice and mixed with 5 μl of isotope standards and incubated for 10 min at 20°C to permit small molecules and vitamins in the internal standards to associate with plasma binding proteins. Macromolecules (protein, DNA, RNA, glycans, etc.) were precipitated by extraction with 4 volumes (400 μl) of cold (-20°C), acetonitrile:methanol (50:50) and incubated on crushed ice for 10 min, then removed by centrifugation at 16,000g x 10 min at 4°C. The supernatants containing the extracted metabolites and internal standards were transferred to labeled cryotubes and stored at -80°C for LCMS/MS analysis. |
Extract Storage: | -80 |
Sample Derivatization: | No |
Sample Spiking: | 25-35 commercial stable isotope internal standards, and 57 stable isotope internal standards that were custom-synthesized in E. coli and S. cerevisiae by metabolic labeling with 13C-glucose and 13C bicarbonate |
Combined analysis:
Analysis ID | AN000946 | AN000947 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | HILIC | HILIC |
Chromatography system | Shimadzu 20AD | Shimadzu 20AD |
Column | Shodex NH2P-40 2E (250× 2.0 mm) | Shodex NH2P-40 2E (250× 2.0 mm) |
MS Type | ESI | ESI |
MS instrument type | Triple quadrupole | Triple quadrupole |
MS instrument name | ABI Sciex 5500 QTrap | ABI Sciex 5500 QTrap |
Ion Mode | NEGATIVE | POSITIVE |
Units | AUC | AUC |
Chromatography:
Chromatography ID: | CH000674 |
Chromatography Summary: | hydrophilic interaction chromatography (HILIC) on shodex column |
Instrument Name: | Shimadzu 20AD |
Column Name: | Shodex NH2P-40 2E (250× 2.0 mm) |
Column Pressure: | 900~2800 psi |
Column Temperature: | 25 C |
Flow Gradient: | 0-3.5 min: 95%B, 3.6-8 min: 85% B, 8.1-13 min: 75% B, 13.5–35 min: 0% B, 36–46 min: 95% B, 46.1 min: end. |
Flow Rate: | 0.2 ml/min |
Injection Temperature: | room temperature |
Solvent A: | 95% water/5% acetonitrile; 20 mM ammonium carbonate; 38 mM ammonium hydroxide, pH 9.8 |
Solvent B: | 100% acetonitrile |
Analytical Time: | 46.1 min |
Oven Temperature: | 25 C |
Sample Loop Size: | 10 ul |
Sample Syringe Size: | 100 ul |
Chromatography Type: | HILIC |
MS:
MS ID: | MS000842 |
Analysis ID: | AN000946 |
Instrument Name: | ABI Sciex 5500 QTrap |
Instrument Type: | Triple quadrupole |
MS Type: | ESI |
Ion Mode: | NEGATIVE |
Collision Energy: | Compound-specific |
Collision Gas: | High |
Ion Source Temperature: | 500 C |
Ion Spray Voltage: | -4500 |
Ionization: | ESI |
Mass Accuracy: | 0.1 Da |
Scanning: | MRM |
MS ID: | MS000843 |
Analysis ID: | AN000947 |
Instrument Name: | ABI Sciex 5500 QTrap |
Instrument Type: | Triple quadrupole |
MS Type: | ESI |
Ion Mode: | POSITIVE |
Collision Energy: | Compound-specific |
Collision Gas: | High |
Ion Source Temperature: | 500 C |
Ion Spray Voltage: | 5500 |
Mass Accuracy: | 0.1 Da |
Scanning: | MRM |