Metadata details for analysis AN000921 | |
---|---|
Study ID | ST000601 |
Analysis ID | AN000921 |
Study Title | Untargeted LC-MS metabolomics analysis of human COPD plasma, HILIC & C18 |
Institute | University of Colorado Denver |
Species | Homo sapiens |
Ion_mode | POSITIVE |
MS type | ESI |
MS Instrument Name | Agilent |
MS Instrument Type | QTOF |
Chromatography Instrument Name | Agilent 6520 |
Chromatography Type | HILIC |
Chromatography Column | Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um) |
Solvent A | 50% acetonitrile/50% water; ammonium acetate, pH 5.8 |
Solvent B | 90% acetonitrile/10% water; pH 5.8 ammonium acetate |
Gradient | 0.2 minutes 100% B, 2-2.1 minutes 100-90% B, 2.1-8.6 minutes 90-50% B, 8.6-8.7 minutes 50-0% B, 8.7-14.7 minutes 0% B, 14.7-14.8 minutes 0-100% B, followed by column re-equilibration |
Flow rate | 0.6 mL/min |
Column Temperature | 20°C |
Retention time units | Minutes |