This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001561. The data can be accessed directly via it's Project DOI: 10.21228/M8XD8F This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST002427 |
Study Title | Dansylated Human Plasma at Known Light-Heavy Mixing Ratios |
Study Type | MS isotopic labeling analysis |
Study Summary | Pooled human plasma was split and tagged with either light or heavy dansyl chloride and then recombined at known light-to-heavy ratios for LC-MS analysis and subsequent testing of isotopic labeling software. |
Institute | University of North Carolina Greensboro |
Last Name | Zhang |
First Name | Qibin |
Address | 600 Laureate Way, Suite 2203, Kannapolis, NC 28081 |
q_zhang2@uncg.edu | |
Phone | +1 (704) 250-5810 |
Submit Date | 2022-12-21 |
Raw Data Available | Yes |
Raw Data File Type(s) | d |
Analysis Type Detail | LC-MS |
Release Date | 2023-01-20 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003950 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Thermo Vanquish |
Column | Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive HF hybrid Orbitrap |
Ion Mode | POSITIVE |
Units | unitless L/H ratio |
Chromatography:
Chromatography ID: | CH002924 |
Chromatography Summary: | A Thermo Scientific Vanquish Horizon UHPLC System coupled with a Thermo Scientific Q Exactive HF Orbitrap was used to conduct LC-(+)FTMS analysis. A Waters ACQUITY UPLC BEH C18 (1.7 μm, 2.1 x 100 mm) column was used along with the following mobile phases: (A) 0.1% formic acid in water and (B) 0.1% formic acid in acetonitrile. The gradient was 25% B (0 min), 25% B (1.2 min), 99% B (11.2 min), 99% B (16.2 min), 25% B (16.3 min), and 25% B (19 min) to achieve column equilibration. The column was maintained at 40 ºC with flowrate of 400 μL/min. |
Instrument Name: | Thermo Vanquish |
Column Name: | Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um) |
Column Temperature: | 40 |
Flow Gradient: | 25% B (0 min), 25% B (1.2 min), 99% B (11.2 min), 99% B (16.2 min), 25% B (16.3 min), 25% B (19 min) |
Flow Rate: | 400 uL/min |
Solvent A: | 100% water, 0.1% formic acid |
Solvent B: | 100% acetonitrile, 0.1% formic acid |
Chromatography Type: | Reversed phase |