Return to study ST001268 main page

MB Sample ID: SA092109

Local Sample ID:GLD054
Subject ID:SU001336
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002105 AN002106 AN002107 AN002108
Analysis type MS MS MS MS
Chromatography type Reversed phase Reversed phase HILIC HILIC
Chromatography system Agilent 1290 Agilent 1290 Agilent 1290 Agilent 1290
Column Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um) Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um) Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um) Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um)
MS Type ESI ESI ESI ESI
MS instrument type QTOF QTOF QTOF QTOF
MS instrument name Agilent 6520 QTOF Agilent 6520 QTOF Agilent 6550 QTOF Agilent 6550 QTOF
Ion Mode POSITIVE NEGATIVE POSITIVE NEGATIVE
Units Abundance Abundance Abundance Abundance

Chromatography:

Chromatography ID:CH001537
Chromatography Summary:Gradient was as follows: 0–1 minutes 30–70% B, 1–7.92 minutes 70–100% B, 7.92–10.4 minutes 100% B, 10.4–10.5 minutes 100–30% B, 10.5–15.1 minutes 30% B.
Instrument Name:Agilent 1290
Column Name:Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um)
Column Temperature:60 °C
Flow Rate:0.7 mL/min
Injection Temperature:4 °C
Solvent A:100% water; 0.1% formic acid
Solvent B:60% isopropanol/36% acetonitrile/4% water; 0.1% formic acid
Chromatography Type:Reversed phase
  
Chromatography ID:CH001538
Chromatography Summary:Gradient elution was as follows: 0–2 minutes 100% B, 2–2.1 minutes 100–90% B, 2.1–8.6 minutes 90–50% B, 8.6–8.7 minutes 50–0% B, 8.7–14.7 minutes 0% B, 14.7–14.8 minutes 0–100% B, 14.8–24.8 minutes 100% B.
Instrument Name:Agilent 1290
Column Name:Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um)
Column Temperature:20 °C
Flow Rate:0.6 mL/min
Injection Temperature:4 °C
Solvent A:50% acetonitrile/50% water; ammonium acetate, pH 5.8
Solvent B:90% acetonitrile/10% water; pH 5.8 ammonium acetate
Chromatography Type:HILIC
  logo