Return to study ST001422 main page

MB Sample ID: SA119154

Local Sample ID:q3June2014_29b_1
Subject ID:SU001496
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:29-82
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002378 AN002379
Analysis type MS MS
Chromatography type HILIC Reversed phase
Chromatography system Dionex UltiMate 3000 Dionex UltiMate 3000
Column Thermo Fisher Accucore HILIC (50 x 2.1mm,2.6um) with Thermo accucore HILIC guard cartridge Thermo Fisher Accucore C18 (50 x 2.1mm,2.6um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE POSITIVE
Units peak area peak area

Chromatography:

Chromatography ID:CH001745
Chromatography Summary:The HILIC column is operated parallel to reverse phase column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6-port switching valves. During operation of HILIC separation method, the MS is operated in positive ion mode and 10 microliters of sample is injected onto the HILIC column while the reverse phase column is flushing with wash solution. Flow rate is maintained at 0.35 mL/min until 2 min, increased to 0.4 mL/min at 6 min and held for 4 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 2% formic acid (v/v) in LC-MS grade water. Initial mobile phase conditions are 25% A, 70% B, 5% C hold for 2 min, with linear gradient to 75% A, 20% B, 5% C at 6 min, hold for 4 min, resulting in a total analytical run time of 10 min. During the flushing phase (reverse phase analytical separation), the HILIC column is equilibrated with a wash solution of 25% A, 70% B, 5% C.
Methods ID:2% formic acid in LC-MS grade water
Methods Filename:20160120_posHILIC120kres10min_ESI_c18poswash_acrore.meth
Chromatography Comments:Triplicate injections for each chromatography mode
Instrument Name:Dionex UltiMate 3000
Column Name:Thermo Fisher Accucore HILIC (50 x 2.1mm,2.6um) with Thermo accucore HILIC guard cartridge
Column Temperature:60C
Sample Injection:10 uL
Solvent A:100% water
Solvent B:100% acetonitrile
Analytical Time:10 min
Sample Loop Size:15 uL
Sample Syringe Size:100 uL
Chromatography Type:HILIC
  
Chromatography ID:CH001746
Chromatography Summary:The C18 column is operated parallel to the HILIC column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6- port switching valves. During operation of the C18 method, the MS is operated in negative ion mode and 10 uL of sample is injected onto the C18 column while the HILIC column is flushing with wash solution. Flow rate is maintained at 0.35 mL/min until 2 min, increased to 0.4 mL/min at 6 min and held for 4 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 2% formic acid (v/v) in LC-MS grade water. Initial mobile phase conditions are 80% A, 15% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 2 min, hold for 4min, resulting in a total analytical run time of 10 min. During the flushing phase (HILIC analytical separation), the C18 column is equilibrated with a wash solution of 0% A, 95% B, 5% C until 7.5 min, followed by an equilibration solution of 80% A, 15% B, 5% C for 2.5 min.
Methods Filename:20160120_posC18120kres10mim_ESI_hilicposwash_acore.meth
Instrument Name:Dionex UltiMate 3000
Column Name:Thermo Fisher Accucore C18 (50 x 2.1mm,2.6um)
Column Temperature:60C
Flow Rate:0.35 mL/min for 2 min; linear increase to 0.4 mL/min at 6 min held for 4 min
Sample Injection:10 uL
Solvent A:100% water
Solvent B:100% acetonitrile
Analytical Time:10 min
Sample Loop Size:15 uL
Sample Syringe Size:100 uL
Chromatography Type:Reversed phase