Return to study ST001717 main page

MB Sample ID: SA161874

Local Sample ID:Mito_007_po
Subject ID:SU001794
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606
Genotype Strain:HeLa cell line
Gender:Not applicable

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002797 AN002798
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Waters Acquity I-Class Waters Acquity I-Class
Column Waters Acquity CSH C18 (100 x 2.1mm,1.7um) Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name ABI Sciex 5600 TripleTOF ABI Sciex 5600 TripleTOF
Ion Mode POSITIVE NEGATIVE
Units Peak area Peak area

Chromatography:

Chromatography ID:CH002067
Chromatography Summary:Chromatographic separation of lipids in cells and mitochondria was performed with an Acquity UPLC system (Waters, USA) using an Acquity UPLC CSH C18 column (2.1100 mm, 1.7 µm; Waters) at 55°C and a flow rate of 0.4 ml/min. The mobile phase for positive ion mode comprised 10 mM ammonium formate in water/acetonitrile (40:60, v/v) containing 0.1% formic acid (solvent A) and isopropanol/acetonitrile (90:10, v/v) containing 0.1% formic acid (solvent B). The mobile phase for negative ion mode comprised 10 mM ammonium acetate in water/acetonitrile (60:40, v/v) (solvent A) and isopropanol/acetonitrile (90:10, v/v) (solvent B). The UPLC gradient was programmed as follows: 40% to 43% B from 0 min to 2 min, 43% to 50% B from 2 min to 2.1 min, 50% to 54% B from 2.1 min to 12 min, 54% to 70% B from 12 min to 12.1 min, 70% to 99% B from 12.1 min to 18 min, 99% to 40% B from 18 min to 18.1 min, and 40% B for 2 min to equilibrate for the next run.
Instrument Name:Waters Acquity I-Class
Column Name:Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
Column Temperature:55
Flow Gradient:40% to 43% B from 0 min to 2 min, 43% to 50% B from 2 min to 2.1 min, 50% to 54% B from 2.1 min to 12 min, 54% to 70% B from 12 min to 12.1 min, 70% to 99% B from 12.1 min to 18 min, 99% to 40% B from 18 min to 18.1 min, and 40% B for 2 min to equilibrate for the next run.
Flow Rate:0.4 ml/min
Solvent A:40% water/60acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Solvent B:90% isopropanol/10% acetonitrile
Chromatography Type:Reversed phase
  logo