Return to study ST002539 main page

MB Sample ID: SA255498

Local Sample ID:Sea ice_T-S_A
Subject ID:SU002639
Subject Type:Water sample

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004181 AN004182 AN004183 AN004184
Analysis type MS MS MS MS
Chromatography type HILIC HILIC Reversed phase Reversed phase
Chromatography system Waters Acquity I-Class Waters Acquity I-Class Waters Acquity I-Class Waters Acquity I-Class
Column SeQuant ZIC- pHILIC (150 x 2.1mm,5um) SeQuant ZIC- pHILIC (150 x 2.1mm,5um) Waters ACQUITY UPLC CSH C18 (150 x 2.1mm,1.7um) Waters ACQUITY UPLC HSS CN (100 x 2.1mm,1.8um)
MS Type ESI ESI ESI ESI
MS instrument type Orbitrap Orbitrap Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE NEGATIVE POSITIVE NEGATIVE
Units Estimated metabolite carbon concentration (nmol C per L) Estimated metabolite carbon concentration (nmol C per L) Estimated metabolite carbon concentration per particulate carbon (nmol C per umol C) Estimated metabolite carbon concentration (nmol C per L)

Chromatography:

Chromatography ID:CH003098
Chromatography Summary:See attached summary.
Instrument Name:Waters Acquity I-Class
Column Name:SeQuant ZIC- pHILIC (150 x 2.1mm,5um)
Column Temperature:30
Flow Gradient:100% B for 2 minutes, ramped to 64% A over 18 minutes, ramped up to 100% A over 1 minute, held at 100% A for 7 minutes, and equilibrated at 100% B for 22 minutes (total time is 50 minutes)
Flow Rate:0.15 mL/min
Solvent A:10mM ammonium carbonate;85% water/15% acetonitrile
Solvent B:10mM ammonium carbonate;85% acetonitrile/15% water
Chromatography Type:HILIC
  
Chromatography ID:CH003099
Chromatography Summary:See attached summary.
Instrument Name:Waters Acquity I-Class
Column Name:Waters ACQUITY UPLC CSH C18 (150 x 2.1mm,1.7um)
Column Temperature:65
Flow Gradient:Initial conditions were 90% A and 10% B. The column was ramped to 80% B over 33 minutes, ramped up to 90% B over 12 minutes, held at 90% B for 1 minute, and equilibrated to 90%A and 10% B for 6 minutes (total time is 52 minutes).
Flow Rate:0.45 mL/min
Solvent A:10mM ammonium formate; 60% acetonitrile/40% water; 0.1% formic acid
Solvent B:10mM ammonium formate; 90% isopropyl alcohol/10% acetonitrile; 0.1% formic acid
Chromatography Type:Reversed phase
  
Chromatography ID:CH003100
Chromatography Summary:See attached summary.
Instrument Name:Waters Acquity I-Class
Column Name:Waters ACQUITY UPLC HSS CN (100 x 2.1mm,1.8um)
Column Temperature:35
Flow Gradient:The column was held at 5% B for 2 minutes, ramped to 100% B over 18 minutes, held at 100% B for 2 minutes, and equilibrated at 5% B for 5 minutes (total run time is 25 minutes)
Flow Rate:0.4 mL/min
Solvent A:0.1% formic acid in water
Solvent B:0.1% formic acid in acetonitrile
Chromatography Type:Reversed phase
  logo