MB Sample ID: SA186768

Local Sample ID:SA184025
Subject ID:SU002076
Subject Type:Bacteria
Subject Species:Escherichia coli
Taxonomy ID:679895
Genotype Strain:BW25113

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003252 AN003253 AN003254
Analysis type MS MS MS
Chromatography type Reversed phase Reversed phase Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 Agilent 1290 Infinity Agilent 1290 Infinity
Column Macherey-Nagel EC 100/2 Nucleoshell C18 2.7μm Waters Acquity BEH C18 (100 x 2mm,1.7um) Waters Acquity BEH C18 (100 x 2mm,1.7um)
MS Type ESI ESI ESI
MS instrument type QTOF QTOF QTOF
MS instrument name Bruker micrOTOFQ II Bruker micrOTOFQ II Bruker maXis II
Ion Mode POSITIVE POSITIVE POSITIVE
Units N/A (structure elucidation) N/A (structure elucidation) N/A (structure elucidation)

MS:

MS ID:MS003024
Analysis ID:AN003252
Instrument Name:Bruker micrOTOFQ II
Instrument Type:QTOF
MS Type:ESI
MS Comments:For standard measurements on the Dionex Ultimate 3000 + Bruker micrOTOFQ II system MS data was acquired over a range from 50 to 2000 m/z in positive mode. Auto MS/MS fragmentation was achieved with rising collision energy (for single charged ions: 18–45 eV over a gradient from 100 to 1000 m/z; for double charged ions: 15–32 eV over a gradient from 100 to 1000 m/z). Calibration of mass spectra was achieved using 10 mM sodium formate in H2O/ iPrOH (1:1) as internal standard. Mass spectra were analysed using Bruker Data Analysis 4.2 software (Bruker Daltonics, Bremen, Germany). Masses of the expected compounds were determined using ChemDraw Professional 16.0.1.4 (PerkinElmer, Waltham, USA) and recorded chromatograms were extracted for the respective m/z. Ions with fitting m/z (Δ<10 ppm) and plausible retention time were fragmented and fragments were matched to in silico identified plausible decay products.
Ion Mode:POSITIVE
Capillary Voltage:3500
Analysis Protocol File:Mass_Spectrometry_Protocol.docx
  
MS ID:MS003025
Analysis ID:AN003253
Instrument Name:Bruker micrOTOFQ II
Instrument Type:QTOF
MS Type:ESI
MS Comments:For standard measurements on the Agilent Infinity 1290 + Bruker micrOTOFQ II system MS data was acquired over a range from 100 to 1500 m/z in positive mode. Auto MS/MS fragmentation was achieved with rising collision energy (for single charged ions: 18–45 eV over a gradient from 100 to 1000 m/z; for double charged ions: 15–32 eV over a gradient from 100 to 1000 m/z). Calibration of mass spectra was achieved using 10 mM sodium formate in H2O/ iPrOH (1:1) as internal standard. Mass spectra were analysed using Bruker Data Analysis 4.2 software (Bruker Daltonics, Bremen, Germany). Masses of the expected compounds were determined using ChemDraw Professional 16.0.1.4 (PerkinElmer, Waltham, USA) and recorded chromatograms were extracted for the respective m/z. Ions with fitting m/z (Δ<10 ppm) and plausible retention time were fragmented and fragments were matched to in silico identified plausible decay products.
Ion Mode:POSITIVE
Capillary Voltage:3500
Analysis Protocol File:Mass_Spectrometry_Protocol.docx
  
MS ID:MS003026
Analysis ID:AN003254
Instrument Name:Bruker maXis II
Instrument Type:QTOF
MS Type:ESI
MS Comments:For standard measurements on the Agilent Infinity 1290 + Bruker maXis II system MS data was acquired over a range from 50 to 2000 m/z in positive mode. Auto MS/MS fragmentation was achieved with rising collision energy (for single charged ions: 35–70 eV over a gradient from 500 to 2000 m/z; for double charged ions: 25–50 eV over a gradient from 500 to 2000 m/z). Calibration of mass spectra was achieved using sodium formate in H2O/ iPrOH (1:1) as internal standard. Mass spectra were analysed using Bruker Data Analysis 4.2 software (Bruker Daltonics, Bremen, Germany). Masses of the expected compounds were determined using ChemDraw Professional 16.0.1.4 (PerkinElmer, Waltham, USA) and recorded chromatograms were extracted for the respective m/z. Ions with fitting m/z (Δ<10 ppm) and plausible retention time were fragmented and fragments were matched to in silico identified plausible decay products.
Ion Mode:POSITIVE
Capillary Voltage:4500
Analysis Protocol File:Mass_Spectrometry_Protocol.docx