Summary of Study ST001415

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000971. The data can be accessed directly via it's Project DOI: 10.21228/M85X14 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  |  Perform analysis on untargeted data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001415
Study TitleMulti-omic profiling of primary mouse neutrophils reveals a pattern of sex and age-related functional regulation
Study SummaryNeutrophils are the most abundant white blood cells in humans and constitute one of the first lines of defense in the innate immune response. Neutrophils are extremely short-lived cells, which survive less than a day after reaching terminal differentiation. Thus, little is known about how organismal aging, rather than the daily cellular aging process, may impact neutrophil biology. In addition, accumulating evidence suggests that both immunity and organismal aging are extremely sex-dimorphic. Here, we describe a multi-omic resource of mouse primary bone marrow neutrophils from young and old female and male animals, at the transcriptomic, metabolomic and lipidomic levels. Importantly, we identify widespread age-related and sex-dimorphic regulation of ‘omics’ in neutrophils, specifically regulation of chromatin metabolism. We leverage machine-learning and identify candidate molecular drivers of age-related and sex-dimorphic transcriptional regulation of neutrophils. We leverage our resource to predict increased levels/release of neutrophil elastase in male mice. To date, this dataset represents the largest multi-omic resource for the study of neutrophils across biological sex and ages. This resource identifies molecular states linked to neutrophil characteristics linked to organismal age or sex, which could be leveraged to improve immune responses across individuals.
Institute
Stanford University
Last NameContrepois
First NameKevin
Address300 Pasteur Dr
Emailkcontrep@stanford.edu
Phone6506664538
Submit Date2020-06-30
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2021-06-30
Release Version1
Kevin Contrepois Kevin Contrepois
https://dx.doi.org/10.21228/M85X14
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002365 AN002366 AN002367 AN002368
Analysis type MS MS MS MS
Chromatography type HILIC HILIC Reversed phase Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 RS Thermo Dionex Ultimate 3000 RS Thermo Dionex Ultimate 3000 RS Thermo Dionex Ultimate 3000 RS
Column SeQuant ZIC-HILIC (100 x 2.1mm,3.5um) SeQuant ZIC-HILIC (100 x 2.1mm,3.5um) Agilent Zorbax SBaq (50 x 2.1mm,1.7um) Agilent Zorbax SBaq (50 x 2.1mm,1.7um)
MS Type ESI ESI ESI ESI
MS instrument type Orbitrap Orbitrap Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Plus Orbitrap Thermo Q Exactive Plus Orbitrap Thermo Q Exactive Plus Orbitrap Thermo Q Exactive Plus Orbitrap
Ion Mode POSITIVE NEGATIVE POSITIVE NEGATIVE
Units MS count MS count MS count MS count

Chromatography:

Chromatography ID:CH001735
Chromatography Summary:HILIC experiments were performed using a ZIC-HILIC column 2.1x100 mm, 3.5μm, 200Å (Merck Millipore) and mobile phase solvents consisting of 10mM ammonium acetate in 50/50 acetonitrile/water (A) and 10 mM ammonium acetate in 95/5 acetonitrile/water (B).(Contrepois et al., 2015)
Instrument Name:Thermo Dionex Ultimate 3000 RS
Column Name:SeQuant ZIC-HILIC (100 x 2.1mm,3.5um)
Column Temperature:40
Flow Rate:0.5 ml/min
Solvent A:95% acetonitrile/5% water; 10 mM ammonium acetate
Solvent B:95% acetonitrile/5% water; 10 mM ammonium acetate
Chromatography Type:HILIC
  
Chromatography ID:CH001736
Chromatography Summary:RPLC experiments were performed using a Zorbax SBaq column 2.1 x 50 mm, 1.7 μm, 100Å (Agilent Technologies) and mobile phase solvents consisting of 0.06% acetic acid in water (A) and 0.06% acetic acid in methanol (B). (Contrepois et al., 2015)
Instrument Name:Thermo Dionex Ultimate 3000 RS
Column Name:Agilent Zorbax SBaq (50 x 2.1mm,1.7um)
Column Temperature:60
Flow Rate:0.6 ml/min
Solvent A:100% water; 0.06% acetic acid
Solvent B:100% methanol; 0.06% acetic acid
Chromatography Type:Reversed phase
  logo