Summary of Study ST001722

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001104. The data can be accessed directly via it's Project DOI: 10.21228/M80M6C This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  |  Perform analysis on untargeted data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001722
Study TitleThe effects of birth weight and breeding value for protein deposition on the plasma metabolome in growing pigs (part-I)
Study SummaryAn experiment was conducted with growing pigs, to determine the effects of birth weight (BiW) and estimated breeding value for protein deposition (EBV) on the metabolomic profile in plasma samples collected under different dietary regimens: protein adequate (A) or protein restricted (R, 70% of A).
Institute
Aarhus University
DepartmentAnimal Science
LaboratoryMetabolomics LC-MS platform Aarhus University Foulum
Last NameHedemann
First NameMette
AddressBlichers Alle 20, Tjele, -, 8830, Denmark
EmailMette.Hedemann@anis.au.dk
Phone51448783
Submit Date2021-03-12
Total Subjects40
Num Males40
Raw Data AvailableYes
Raw Data File Type(s)mzXML
Analysis Type DetailLC-MS
Release Date2021-03-31
Release Version1
Mette Hedemann Mette Hedemann
https://dx.doi.org/10.21228/M80M6C
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002806
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 RS
Column Waters Acquity BEH HSS T3 (100 x 2.1mm,1.8um)
MS Type ESI
MS instrument type QTOF
MS instrument name Bruker Impact HD
Ion Mode NEGATIVE
Units peak area

Chromatography:

Chromatography ID:CH002073
Chromatography Summary:Chromatographic separation was performed on a Dionex UltiMate 3000RSL Binary UHPLC System (Thermo Scientific Dionex, Sunnyvale, CA) equipped with a HSS T3 C18 UHPLC column, 1.8 µm, 100x2.1 mm (Waters Corporation, Milford, MA). The column was maintained at 30°C. Samples were kept in the autosampler at 10°C, and the injection volume for both sample types was 3 µl. The mobile phases were 0.1% formic acid in Milli-Q water (A) and 0.1% formic acid in acetonitrile (B). The gradient program for the plasma samples was as follows: 0-12 min, linear gradient from 5-100% B; 12-13 min, 100% B and return to initial conditions in 0.2 min. In all cases, the column was re-equilibrated at 5% B for 2 min at the beginning of each run. The flow rate was 0.4 ml/min
Instrument Name:Thermo Dionex Ultimate 3000 RS
Column Name:Waters Acquity BEH HSS T3 (100 x 2.1mm,1.8um)
Column Temperature:30
Flow Rate:0.4 ml/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:Reversed phase