Summary of Study ST003159

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001964. The data can be accessed directly via it's Project DOI: 10.21228/M8VD96 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files
Study IDST003159
Study TitleUntargeted serum metabolomics in the Parkinson's Environment and Genes (PEG) Study
Study SummaryThis project aims to evaluate the serum metabolome of Parkinson’s disease (PD) patients relative to unaffected controls in the Parkinson’s Environment and Genes (PEG) Study. Background: Untargeted high-resolution metabolomic profiling provides simultaneous measurement of thousands of metabolites. Metabolic networks based on these data can help uncover disease-related perturbations across interconnected pathways. Objective: Identify metabolic disturbances associated with PD in the PEG population-based study using untargeted metabolomics. Methods: We provide serum-based untargeted metabolomics data derived from liquid chromatography with high-resolution mass spectrometry (LC-HRMS). LC-HRMS detected 4,762 metabolites for analysis (HILIC: 2716 metabolites; C18: 2046 metabolites).
Institute
University of California, Los Angeles
DepartmentNeurology
Last NamePaul
First NameKimberly
Address710 Westwood Plaza, Los Angeles, CA, 90095, USA
Emailkimberlp@ucla.edu
Phone310.206.7458
Submit Date2023-12-15
Total Subjects919
Analysis Type DetailLC-MS
Release Date2024-04-23
Release Version1
Kimberly Paul Kimberly Paul
https://dx.doi.org/10.21228/M8VD96
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN005181 AN005182
Analysis type MS MS
Chromatography type HILIC Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 Thermo Dionex Ultimate 3000
Column Waters XBridge BEH Amide (50 x 2.1mm,2.5um) Thermo Higgins C18 (50 x 2.1mm,3um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE NEGATIVE
Units log2 transformed intensity log2 transformed intensity

Chromatography:

Chromatography ID:CH003919
Chromatography Summary:The HILIC column is operated parallel to reverse phase column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6-port switching valves. During operation of HILIC separation method, the MS is operated in positive ion mode and 10 microliters of sample is injected onto the HILIC column while the reverse phase column is flushing with wash solution. Flow rate is maintained at 0.35 mL/min until 1.5 min, increased to 0.4 mL/min at 4 min and held for 1 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 2% formic acid (v/v) in LC-MS grade water. Initial mobile phase conditions are 22.5% A, 75% B, 2.5% C hold for 1.5 min, with linear gradient to 77.5% A, 20% B, 2.5% C at 4 min, hold for 1 min, resulting in a total analytical run time of 5 min. During the flushing phase (reverse phase analytical separation), the HILIC column is equilibrated with a wash solution of 77.5% A, 20% B, 2.5% C.
Methods Filename:EmoryUniversity_HRM-QEHF_chromatography_5min_092017_v1.pdf
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Waters XBridge BEH Amide (50 x 2.1mm,2.5um)
Column Temperature:60
Flow Gradient:22.5% A, 75% B, 2.5% C hold for 1.5 min, linear gradient to 77.5% A, 20% B, 2.5% C at 4 min, hold for 1 min
Flow Rate:0.35 mL/min for 1.5 min; linear increase to 0.4 mL/min at 4 min, hold for 1 mi
Sample Injection:10 uL
Solvent A:100% water
Solvent B:100% acetonitrile
Analytical Time:5 min
Chromatography Type:HILIC
Solvent C:98% water/2% formic acid
  
Chromatography ID:CH003920
Chromatography Summary:The C18 column is operated parallel to the HILIC column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6- port switching valves. During operation of the C18 method, the MS is operated in negative ion mode and 10 μL of sample is injected onto the C18 column while the HILIC column is flushing with wash solution. Flow rate is maintained at 0.4 mL/min until 1.5 min, increased to 0.5 mL/min at 2 min and held for 3 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 10mM ammonium acetate in LC-MS grade water. Initial mmobile phase conditions are 60% A, 35% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 1.5 min, hold for 3.5 min, resulting in a total analytical run time of 5 min. During the flushing phase (HILIC analytical separation), the C18 column is equilibrated with a wash solution of 0% A, 95% B, 5% C until 2.5 min, followed by an equilibration solution of 60% A, 35% B, 5% C for 2.5 min.
Methods Filename:EmoryUniversity_HRM-QEHF_chromatography_5min_092017_v1.pdf
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Thermo Higgins C18 (50 x 2.1mm,3um)
Column Temperature:60
Flow Gradient:60% A, 35% B, 5% C hold for 0.5 min, linear gradient to 0% A, 95% B, 5% C at 1.5 min, hold for 3 min
Flow Rate:0.4 mL/min for 1.5 min; linear increase to 0.5 mL/min at 2 min held for 3 min
Sample Injection:10 uL
Solvent A:100% water
Solvent B:100% acetonitrile
Analytical Time:5 min
Chromatography Type:Reversed phase
Solvent C:100% water; 10mM ammonium acetate