This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001049. The data can be accessed directly via it's Project DOI: 10.21228/M83Q4K This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST001640 |
Study Title | Remodeling Lipids in the Transition from Chronic Liver Disease to Hepatocellular Carcinoma (Blood) - part I |
Study Summary | Comparing blood lipidomics of healthy volunteers to patients with chronic liver disease (CLD), and to patients with HCC caused by viral infections. We contrasted our findings in blood to lipid alterations in liver tumor and nontumor tissue samples from HCC patients. |
Institute | University of California, Davis |
Last Name | Ismail |
First Name | Israa |
Address | 451 Health Sciences Drive, Room 1313, Davis, CA, 95616, USA |
Israataher2015@gmail.com | |
Phone | 01 530 7613155 |
Submit Date | 2021-01-04 |
Raw Data Available | Yes |
Raw Data File Type(s) | d |
Analysis Type Detail | GC-MS |
Release Date | 2021-01-25 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN002685 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Agilent |
Column | Waters ACQUITY UPLC CSH C18 |
MS Type | EI |
MS instrument type | QTOF |
MS instrument name | Agilent 6530 QTOF |
Ion Mode | POSITIVE |
Units | counts per second |
MS:
MS ID: | MS002484 |
Analysis ID: | AN002685 |
Instrument Name: | Agilent 6530 QTOF |
Instrument Type: | QTOF |
MS Type: | EI |
MS Comments: | The quadrupole/time-of-flight (QTOF) mass spectrometers are operated with electrospray ionization (ESI) performing full scan in the mass range m/z 65–1700 in positive (Agilent 6530, equipped with a JetStreamSource) and negative (Agilent 6550, equipped with a dual JetStream Source) modes producing both unique and complementary spectra. Instrument parameters are as follows (positive mode) Gas Temp 325°C, Gas Flow 8 l/min, Nebulizer 35 psig, Sheath Gas 350°C, Sheath Gas Flow 11, Capillary Voltage 3500 V, Nozzle Voltage 1000V, Fragmentor 120V, Skimmer 65V. Data (both profile and centroid) are collected at a rate of 2 scans per second. In negative ion mode, Gas Temp 200°C, Gas Flow 14 l/min, Fragmentor 175V, with the other parameters identical to positive ion mode. For the 6530 QTOF, a reference solution generating ions of 121.050 and 922.007 m/z in positive mode and 119.036 and 966.0007 m/z in negative mode, and these are used for continuous mass correction. For the 6550, the reference solution is introduced via a dual spray ESI, with the same ions and continuous mass correction. Samples are injected (1.7 μl in positive mode and 5 μl in negative ion mode) with a needle wash for 20 seconds (wash solvent is isopropanol). The valve is switched back and forth during the run for washing; this has been shown to be essential for reducing carryover of less polar lipids. |
Ion Mode: | POSITIVE |