Summary of Study ST002900

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001773. The data can be accessed directly via it's Project DOI: 10.21228/M8J420 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002900
Study TitleFolate metabolite levels in erythroid progenitor cells following SHIN1 and inosine treatment
Study SummaryCulture of erythroid progenitor cells cultured in SFEM II media supplemented with SHIN1 and/or inosine for 48 hours followed up by metabolomics targeting folate metabolites.
Institute
Boston Children's Hospital, Harvard Medical School
Departmentpathology
LaboratoryKanarek Lab
Last NameKanarek
First NameNaama
AddressEnders 1116.2, 300 Longwood Ave, Boston, MA 02115
Emailnaama.kanarek@childrens.harvard.edu
Phone6173557433
Submit Date2023-10-01
Num Groups4
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2024-02-13
Release Version1
Naama Kanarek Naama Kanarek
https://dx.doi.org/10.21228/M8J420
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004759
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Vanquish
Column Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode UNSPECIFIED
Units peak area

MS:

MS ID:MS004505
Analysis ID:AN004759
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The mass spectrometer was operated in full-scan, positive ionization mode using three narrow-range scans: 438–450 m/z; 452–462 m/z; and 470–478 m/z, with the resolution set at 70,000, the AGC target at 10e6, and the maximum injection time of 150 ms. HESI settings were: sheath gas flow rate: 40; Aux gas flow rate: 10; Sweep gas: 0; Spray voltage: 2.8 (neg) 3.5 (pos); Capillary temperature 300; S-lens RF level 50; Aux gas heater temp: 350.
Ion Mode:UNSPECIFIED