Summary of Study ST001217

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000816. The data can be accessed directly via it's Project DOI: 10.21228/M86Q4H This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001217
Study TitleLCMS untargeted Plasma analysis from COPD subjects
Study SummaryTo use LCMS untargeted metabolomics for the purpose of detecting metabolites in 115 matched BAL and plasma that are associated with COPD
Institute
University of Colorado Anschutz Medical Campus; National Jewish Health
Last NameReisdorph
First NameNichole
Address12850 E Montview Blvd
EmailNichole.Reisdorph@cuanschutz.edu
Phone303-720-9234
Submit Date2019-07-10
Raw Data AvailableYes
Raw Data File Type(s)d
Analysis Type DetailLC-MS
Release Date2020-07-15
Release Version1
Nichole Reisdorph Nichole Reisdorph
https://dx.doi.org/10.21228/M86Q4H
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002029 AN002030
Analysis type MS MS
Chromatography type Reversed phase HILIC
Chromatography system Agilent 6545 Agilent 6520
Column Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um) Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Agilent 6545 QTOF Agilent 6520 QTOF
Ion Mode POSITIVE POSITIVE
Units Abundance (Log2) Abundance (Log2)

Chromatography:

Chromatography ID:CH001470
Chromatography Summary:Reversed phase samples from the lipid fraction were randomized in the worklist and run randomly in triplicate using an Agilent 1290 series pump with an Agilent Zorbax Rapid Resolution HD (RRHD) SB-C18, 1.8 micron (2.1 × 100 mm) analytical column and an Agilent Zorbax SB-C18, 1.8 micron (2.1 × 5 mm) guard column. The autosampler tray temperature was set at 4 °C, column temperature was set at 60 °C, and the sample injection volume was 8 µL for BAL and 4 µL for plasma. The flow rate was 0.7 mL/min with the following mobile phases: mobile phase A was water with 0.1% formic acid, and mobile phase B was 60:36:4 isopropyl alcohol:acetonitrile:water with 0.1% formic acid. Gradient elution was as follows: 0–0.5 minutes 30–70% B, 0.5–7.42 minutes 70–100% B, 7.42–10.4 minutes 100% B, 10.4–10.5 minutes 100–30% B, 10.5–15.1 minutes 30% B.
Instrument Name:Agilent 6545
Column Name:Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um)
Column Temperature:60
Flow Gradient:0-0.5 minutes 30-70% B, 0.5-7.42 minutes 70-100% B, 7.42-10.4 minutes 100% B, 10.4-10.5 minutes 100-30% B, 10.5-15.1 minutes 30% B
Flow Rate:0.7 mL/min
Solvent A:100% water; 0.1% formic acid,
Solvent B:60% isopropanol/36% acetonitrile/4% water; 0.1% formic acid
Chromatography Type:Reversed phase
  
Chromatography ID:CH001471
Chromatography Summary:The samples from the aqueous small molecule fraction were analyzed randomly in triplicate using an Agilent 1290 series pump using a Phenomenex Kinetex HILIC, 2.6 µm, 100 Å (2.1 × 50 mm) analytical column and an Agilent Zorbax Eclipse Plus-C8 5 µm (2.1 × 12.5 mm) narrow bore guard column. The autosampler tray temperature was set at 4 °C, column temperature was set at 20 °C, and the sample injection volume was 1 µL for both BAL and plasma. The flow rate of 0.6 mL/min with the following mobile phases: mobile phase A was 50% ACN with pH 5.8 ammonium acetate, and mobile phase B was 90% ACN with pH 5.8 ammonium acetate. Gradient elution was as follows: 0.2 minutes 100% B, 0.2–2.1 minutes 100–90% B, 2.1–8.6 minutes 90–50% B, 8.6–8.7 minutes 50–0% B, 8.7–14.7 minutes 0% B, 14.7–14.8 minutes 0–100% B, 14.8–24.8 minutes 100% B.
Instrument Name:Agilent 6520
Column Name:Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um)
Column Temperature:20
Flow Gradient:0.2 minutes 100% B, 0.2-2.1 minutes 100-90% B, 2.1-8.6 minutes 90-50% B, 8.6-8.7 minutes 50-0% B, 8.7-14.7 minutes 0% B, 14.7-14.8 minutes 0-100% B, 14.8-24.8 minutes 100% B.
Flow Rate:0.6 mL/min
Solvent A:50% acetonitrile/50% water; ammonium acetate, pH 5.8
Solvent B:90% acetonitrile/5% water, pH 5.8
Chromatography Type:HILIC