Summary of Study ST002383

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001533. The data can be accessed directly via it's Project DOI: 10.21228/M8J99C This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002383
Study TitleMetabolomics of B-cell Acute Lymphoblastic Leukemia in Response to Adipocyte Conditioned Media
Study TypeUntargeted MS
Study SummaryAdipocyte conditioned media (ACM), stromal cell conditioned media (SCM) and unconditioned media (UCM) were added to B-cell Acute Lymphoblastic Leukemia cells (REH and RCH-AcV) either with or without methotrexate (MTX). The metabolomic profiles of the cells was determined by mass spectrometry.
Institute
Emory University
DepartmentPediatrics
LaboratoryJoshua Chandler, PhD
Last NameChandler
First NameJoshua
Address2015 Uppergate Drive NE, Atlanta, GA 30322
Emailjoshua.chandler@emory.edu
Phone404-727-3536
Submit Date2022-10-07
Num GroupsTwo sets of two sets of three groups (MTX and no-MTX, REH and RCH-AcV, ACM vs SCM vs UCM)
Total Subjects3 replicates of each group (36 total)
Publicationssubmitted to JNCI
Raw Data AvailableYes
Raw Data File Type(s)mzML
Analysis Type DetailLC-MS
Release Date2023-06-07
Release Version1
Joshua Chandler Joshua Chandler
https://dx.doi.org/10.21228/M8J99C
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003882
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Vanquish
Column SeQuant ZIC-HILIC (100 x 2.1mm,3.5um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE
Units normalized peak area

Chromatography:

Chromatography ID:CH002876
Chromatography Summary:A VanquishTM Horizon Binary ultrahigh performance liquid chromatography system coupled to a Q Exactive High Field Hybrid Orbitrap mass spectrometer (ThermoFisher) was used for data collection. A SeQuant® ZIC-HILIC column (Millipore Sigma, 3.5 µm, 100 Å, 150 x 2.1 mm, PEEK) was pumped with a mixture of mobile phases of 0.1% formic acid in water (mobile phase A) or 0.1% formic acid in acetonitrile (mobile phase B) (all reagents from Millipore Sigma). The column was equilibrated with 90% B for 8.3 min at 0.25 mL/min, after which 2.5 mL of extracted sample was injected. After this, a gradient ran from 90% to 20% B over 17.5, then was held at 20% B at 0.35 mL/min for 7.5 min.
Instrument Name:Thermo Vanquish
Column Name:SeQuant ZIC-HILIC (100 x 2.1mm,3.5um)
Flow Gradient:he column was equilibrated with 90% B for 8.3 min at 0.25 mL/min, after which 2.5 mL of extracted sample was injected. After this, a gradient ran from 90% to 20% B over 17.5 min, then was held at 20% B at 0.35 mL/min for 7.5 min.
Flow Rate:0.25 mL/min-0.35 mL/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:HILIC