Summary of Study ST000366
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000291. The data can be accessed directly via it's Project DOI: 10.21228/M8F600 This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
Study ID | ST000366 |
Study Title | Metabolomics analysis of colon adenoma in African Americans |
Study Type | Metabolomics |
Study Summary | This NMR Metabolomics analysis was performed on feces samples derived from healthy (n = 10) and adenoma (n = 10) African American subjects with the goal of identifying perturbations in metabolomics profiles in colon cancer. |
Institute | University of North Carolina |
Department | Systems and Translational Sciences |
Laboratory | Sumner Lab |
Last Name | Sumner |
First Name | Susan |
Address | Eastern Regional Comprehensive Metabolomics Resource Core, UNC Nutrition Research Institute, 500 Laureate Way, Kannapolis, NC, 28081 |
susan_sumner @unc.edu | |
Phone | 704-250-5066 |
Num Groups | 2 |
Total Subjects | 20 |
Raw Data Available | Yes |
Raw Data File Type(s) | fid |
Analysis Type Detail | NMR |
Release Date | 2017-07-10 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
NMR:
NMR ID: | NM000066 |
Analysis ID: | AN000599 |
Instrument Name: | Bruker950 MHz |
Instrument Type: | Other |
NMR Experiment Type: | Other |
NMR Comments: | nmr experiment type: noesypr1d |
Field Frequency Lock: | Deuterium |
Standard Concentration: | 0.5 mM |
Spectrometer Frequency: | 950 MHz |
NMR Probe: | 5 mm ATMA Cryoprobe |
NMR Solvent: | D2O |
NMR Tube Size: | 5mm, 7 inch |
Shimming Method: | Topshim |
Pulse Sequence: | noesypr1d |
Water Suppression: | yes |
Pulse Width: | 8.44 us |
Power Level: | 12.589 W |
Receiver Gain: | 4 |
Offset Frequency: | 4471.1 Hz |
Presaturation Power Level: | 0.00003588 W |
Chemical Shift Ref Cpd: | DSS |
Temperature: | 298.1 K |
Number Of Scans: | 256 |
Dummy Scans: | 4 |
Acquisition Time: | 1.075 s |
Relaxation Delay: | 2 s |
Spectral Width: | 16 ppm, 15.243 kHz |
Num Data Points Acquired: | 32k |
Real Data Points: | 32k |
Line Broadening: | 0.5 Hz |
Zero Filling: | yes |
Apodization: | Lorentzian |
Baseline Correction Method: | Polynomial |
Chemical Shift Ref Std: | DSS |
Binned Increment: | 0.04 |
Binned Data Excluded Range: | 4.68 - 4.88 (water) |